Exploring Accuracy and Uncertainty in Computed Tomography
2月 15, 2021 | Dirk Steiner
There is a range of terminology for CT system accuracy and uncertainty that should be understood when evaluating 3D scanning systems, this blog post illustrates terms, such as Precision and Accuracy, Tolerance, Uncertainty of Measurement.
Fig. 1: Precision and Accuracy
Precision and Accuracy:
For understanding the difference between accuracy and precision, consider a target like on a dartboard, illustrated in figure 1.
On the upper left the shots are all over the place and relative to the lower right, the precision and accuracy are both low. The top right shows what is sometimes called “good grouping,” all of the shots are very close to each other but not very close to the center which is the aiming point showing low accuracy but high precision. On the lower left all shots are around the center some are just scattered inside the second ring. That would be considered high accuracy but low precision.
Tolerance:
The term tolerance refers to the difference between the upper (maximum) limit and lower (minimum) limit of a dimension. In other words, tolerance is the maximum permissible variation in a dimension.
Uncertainty of Measurement:
In most cases when people are asking about accuracy, what they really mean is the uncertainty of measurement. Although a slightly clumsy term, “uncertainty of measurement” is the phrasing used by many standards that apply to CT scanning.
Fig. 2: Uncertainty and Tolerance
Consider figure 2, the measurement came out with a reading of V1. The uncertainty of measurement is a value of how much the real value is likely to be at a certain confidence level.
Resolution:
The important parameter to know when evaluating a CT dataset is the smallest piece of information in it: voxel size. It is simply defined by the length (or diameter) of the field of view divided by number of voxels created in one direction.
Fig. 3: CT System Components and Field of View
The voxel size doesn’t say anything about the smallest detail visible. If there is a pore the size of one voxel, that it would not be enough information to reliably detect it. A rule of thumb is that at least 2 x 2 x 2 voxels are needed to be able to see a single feature. For the purpose of metrology, that is not enough “meet” to create an accurate surface around the object. The term “resolution” is defined for the purpose of the specific application.
Error:
Error describes the CT system performance and is the deviation from a known measurement under specific conditions. A system is specified to perform maximum permissible error (MPE); there may be multiple different specifications for one system. The error is not the same as uncertainty of measurement. It represents a value that can be checked as part of an acceptance procedure or to monitor the system while in use. The error is measured under ideal conditions defined by the manufacturer often expressed as MPEXE = V μm * (L / D mm)
Suitability:
Is the process suitable to use for a specific measurement task? Answer is black and white: yes or no.
Suitability typically also covers the question in situations as shown above. The measured value V1 is right inside the tolerance even with worst case uncertainty U considered. V2 is outside even if the best-case uncertainty is assumed. But what happens if the value V2 is within the tolerance, but the uncertainty could make the part fail? One approach is to make the tolerance tighter, but a lot of good parts might fail. The way commonly used standards deal with it is to perform a suitability study. Multiple approaches are available e.g. Gauge R&R, MSA, VDA5, VDI/VDE 2630. The study typically includes the entire process including operator and the object to be measured.
最新の投稿
学生インターンが開発?CT画像で砂糖と塩を見分けるAI
2月 12, 2026 | Dr. Daniel Stickler
Comet Yxlonで実施された3週間の学生インターンシップで、私は好奇心旺盛で意欲的な学生を指導する機会を得ました。ただし、安全上の理由から、彼女がX線装置を単独で操作することはできません。 そこで私たちは、実際の業務に近い形で取り組めるプロジェクトを用意しました。それは実データを使った解析や明確な成果目標を設定し、実践的な体験ができる内容です。X線イメージングやCTの基礎を学びながら、ソフトウェアを使ったデータ解析を中心に、自ら手を動かして取り組める内容にしました。
もっと見る225 kV マイクロフォーカス管向け新 HDP ターゲット登場。高出力密度で高速・高分解能検査を実現。
11月 18, 2025
Comet Yxlon の FF35 CT および FF85 CT は、デュアルチューブ設計による高い柔軟性と卓越した精度が最大の特徴です。半導体パッケージ、アディティブマニュファクチャリング、自動車部品、材料研究など、多岐にわたる分野で活用されており、内部構造を迅速かつ高精度に可視化することが求められる現場においてその実力を発揮します。 そして今回、これらのシステムがさらなる進化を遂げました。両機種に搭載可能な 225 kV マイクロフォーカス・ディレクショナルビーム管に、新たに高出力密度(HDP)ターゲットを採用。従来と同じ筐体サイズおよび水冷式焦点安定性を維持しながら、出力密度を2倍に向上させました。 これにより、検査時間の短縮、より精細な画像取得、そして長期にわたる安定したパフォーマンスを実現します。
もっと見るテンの頭蓋骨を高精細CTで可視化
10月 05, 2023 | Gina Naujokat
Comet Yxlon FF35 CT検査システムと高度なVista機能を活用すれば、必要な分解能を選ぶだけで、簡単かつ迅速にスキャン設定が行えます。貴重な時間を節約しながら、高品質なスキャン結果を取得できます。その仕組みをぜひご覧ください。
もっと見る
Accuracy of computed tomography (pdf)