Understanding Guidelines for System Performance

4月 26, 2021 | Dirk Steiner

The gold standard for 3D measurement equipment is the DIN EN ISO 10360 Geometrical Product Specification (GPS). It is designed to help compare different equipment and create a clear understanding of when a system is within the specification and when

The gold standard for 3D measurement equipment is the DIN EN ISO 10360 Geometrical product specification (GPS). It is designed to help compare different equipment and create a clear understanding of when a system is within the specification and when it is not. There two series of standards under the VDI/VDE umbrella that are applicable for CT:

-VDI/VDE 2617 Accuracy of coordinate measuring machines
-VDI/VDE 2630 Computed tomography in dimensional measurement

These crossover standards offer the guideline for the application of DIN EN ISO 10360 for coordinate measurement machines with CT-sensors. VDI/VDE 2630 part 1.3 provides three guidelines:

1.Fundamentals: Definitions of characteristics (errors of probing, length measurement, size, and form) and limits
2.Acceptance test: How to determine a CT system meets the limits of the characteristics
3.Monitoring: Long-term compliance with the characteristics

The standard recommends the use of spheres or spherical caps made of a “suitable material”. A common material is ruby since high-quality ruby spheres are commercially available and are thermally stable. The spheres by themselves allow for determining probing error form and size. The results are compared to the manufacturer’s limits and a compliance report can be created.

A common way to decide if the process is suitable for the application is to compare the tolerance and uncertainty. Here is a list of common standards used in the industry:

-VDI/VDE 2630 Computed tomography in dimensional measurement, Part 2.1
-Germany: VDA 5
-USA: Measurement Systems Analysis (MSA, guide: ASTM E2782)
-Gage repeatability and reproducibility (GR&R, standard practice: ASTM E177)
-Company specific standards (mostly based on MSA: e.g. GM EMS, Ford EU 1880)

Fig.1: Ruby sphere specimen

Part 2.1 of the VDI/VDE 2630 was developed for users of industrial CT systems. Part 1.3 was written so buyer and seller can agree when and how a CT system is performing to meet the specified characteristics. The specified MPEs (maximum permissible error) have nothing to do with a specific application. That’s where part 2.1 comes in play. It serves two purposes:

1.To determine the uncertainty of measurement of specific features
2.To give the user help deciding if the process is suitable to be used for the required tolerance of that feature

It is based on GUM and traceable. The core to determine is this formula for the expanded uncertainty of measurement.

Fig.2: Equation for the Uncertainty of Measurement

U: Expanded uncertainty of measurement
k: coverage factor (2 for 95%)
ucal: uncertainty of the calibration measurement
up: standard uncertainty of the process
uw: uncertainty of the workpiece
b: systematic deviation

To solve the equation several measurements must be run. The standard suggests having one sample calibrated with a method that is more accurate, normally done in a certified lab with a highly accurate CMM. The certified lab will provide the uncertainty of the measurement (ucal).

Download the complete white paper free of charge.

close.png

最新の投稿

学生インターンが開発?CT画像で砂糖と塩を見分けるAI

2月 12, 2026 | Dr. Daniel Stickler

Comet Yxlonで実施された3週間の学生インターンシップで、私は好奇心旺盛で意欲的な学生を指導する機会を得ました。ただし、安全上の理由から、彼女がX線装置を単独で操作することはできません。 そこで私たちは、実際の業務に近い形で取り組めるプロジェクトを用意しました。それは実データを使った解析や明確な成果目標を設定し、実践的な体験ができる内容です。X線イメージングやCTの基礎を学びながら、ソフトウェアを使ったデータ解析を中心に、自ら手を動かして取り組める内容にしました。

もっと見る

225 kV マイクロフォーカス管向け新 HDP ターゲット登場。高出力密度で高速・高分解能検査を実現。

11月 18, 2025

Comet Yxlon の FF35 CT および FF85 CT は、デュアルチューブ設計による高い柔軟性と卓越した精度が最大の特徴です。半導体パッケージ、アディティブマニュファクチャリング、自動車部品、材料研究など、多岐にわたる分野で活用されており、内部構造を迅速かつ高精度に可視化することが求められる現場においてその実力を発揮します。 そして今回、これらのシステムがさらなる進化を遂げました。両機種に搭載可能な 225 kV マイクロフォーカス・ディレクショナルビーム管に、新たに高出力密度(HDP)ターゲットを採用。従来と同じ筐体サイズおよび水冷式焦点安定性を維持しながら、出力密度を2倍に向上させました。 これにより、検査時間の短縮、より精細な画像取得、そして長期にわたる安定したパフォーマンスを実現します。

もっと見る

テンの頭蓋骨を高精細CTで可視化

10月 05, 2023 | Gina Naujokat

Comet Yxlon FF35 CT検査システムと高度なVista機能を活用すれば、必要な分解能を選ぶだけで、簡単かつ迅速にスキャン設定が行えます。貴重な時間を節約しながら、高品質なスキャン結果を取得できます。その仕組みをぜひご覧ください。

もっと見る